P p f p, Zero ng, Cal g 1 – RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 58

Advertising
background image

53 • Calibrating a GX-2012

If one or more of the sensors fails the fresh air adjustment, the SDM-2012
will abort the calibration and will not apply calibration gas. If this happens,
the CAL

S

LED will flash red indicating a failure and the following screens

will appear indicating which channels passed and failed the fresh air
adjustment with a P (pass) or an F (fail). In the following examples, the CO
channel failed the fresh air adjustment.

Figure 27: Failed Fresh Air Adjustment, Standard 4-Gas GX-2012 or GAS Menu

Item in Edit Mode Set to Std

Figure 28: Failed Fresh Air Adjustment, %Volume Sensor Installed and GAS Menu

Item in Edit Mode Set to ALL

In this case continue with step 19.

15. The SDM-2012 will apply GAS 1 calibration gas to the instrument for the time

defined by the GAS TIME calibration parameter.

16. If your GX-2012 does not have a %volume sensor installed, or if the GAS

P
P
F
P

%LEL

%

ppm

ppm

CH

4

O

2

CO

H

2

S

ZERO NG

0
20.9
90
0.0

%LEL

%

ppm

ppm

CH

4

O

2

CO

H

2

S

ZERO NG



P
P
F
P

%

ppm

ppm

CH

4

O

2

CO

H

2

S

ZERO NG

vol%

%LEL

0
20.9
90
0.0

%

ppm

ppm

CH

4

O

2

CO

H

2

S

ZERO NG

vol%

%LEL

45
10.0
47
23.0

%LEL

%

ppm

ppm

CH

4

O

2

CAL G 1

CO

H

2

S