P p f p, Zero ng, Bump g 1 – RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 45

Advertising
background image

Bump Testing a GX-2012 • 40

If one or more of the sensors fails the fresh air adjustment, the SDM-2012
will abort the bump test and will not apply calibration gas. If this happens,
the BUMP

T

LED will flash red indicating a failure and the following screen

will appear indicating which channels passed and failed the fresh air
adjustment with a P (pass) or an F (fail). In the following examples, the CO
channel failed.

Figure 16: Failed Fresh Air Adjustment, Standard 4-Gas GX-2012 or GAS Menu

Item in Edit Mode Set to Std

Figure 17: Failed Fresh Air Adjustment, %Volume Sensor Installed and GAS Menu

Item in Edit Mode Set to ALL

In this case, continue with step 22.

16. The SDM-2012 will apply GAS 1 calibration gas to the instrument for the time

defined by the GAS TIME bump test parameter.