Bump Testing a Gas Tracer • 62

14. Remove any connected cylinders and press and release the EDIT ENTER

button. The SDM-2012 will purge for 30 seconds.

15. At the end of the purge cycle, you will be prompted to reconnect the calibration

cylinder to the GAS 2 fitting.

16. Reconnect the ppm calibration cylinder and press and release the EDIT

ENTER button.

17. The SDM-2012 begins the bump test by applying fresh air to the ppm sensor

for the time defined by the AIR TIME bump test parameter.

NOTE:

If the GAS menu item in Edit Mode is set to Std, the ppm sensor and
%volume sensor will not be bump tested.

18. The SDM-2012 will perform a fresh air adjustment on the ppm sensor.

If the air adjustment is successful, the SDM-2012 will continue as
described in step 19 below.




PURGE

GAS2
On

PURGE

0


ppm

CH

4

BUMP AIR