E air--in, Decision, Transmit – RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 28: Setting the calibration parameters

Advertising
background image

23 • Setting the Operational Parameters in Edit Mode

6.

To edit a parameter, press and release the EDIT ENTER button when that
parameter is displayed. An “E” will appear to the left of the parameter name.

7.

Use the BUMP

T

and CAL

S

buttons to set the parameter to the desired

value, then press and release the EDIT ENTER button. The “E” to the right of
parameter name will disappear.

8.

Repeat step 6 - step 7 to set any other parameters.

9.

When you are done setting the parameters, use the BUMP

T

button to scroll

past the AUTO--C parameter screen. The screen will indicate that the
parameter changes have been saved.

The instrument will return to the home screen.

Setting the Calibration Parameters

Do the following to set the calibration parameters.

1.

Turn on the SDM-2012 with an instrument and establish a connection between
them as described above in "Turning on the SDM-2012 with an Instrument".
The instrument will display the home screen.

30


E AIR--IN

30
30
50
OFF

DECISION

50
12.0
50
25.0

%LEL

%

ppm

ppm

CH

4

O

2

TRANSMIT

CO

H

2

S

Advertising