85 • Calibrating a Gas Tracer

27. After the fresh air purge is complete, the screen will alternate between the gas

readings and the calibration results. The calibration results are indicated with a
P (pass) or an F (fail) to the left of the channel units. If the calibration passed,
the auto calibration values will be displayed. If the calibration failed, the gas
readings at the end of the gas application will be displayed. Use the EDIT
ENTER button to scroll between the calibration results for the standard three
sensors and the calibration results for the ppm sensor and %volume sensor (if
installed). The CAL

S

LED will stop blinking and be steadily green if all

channels passed the calibration or steadily red if any channel failed the
calibration. In the example below, both the CO and ppm sensors failed the
calibration.

Figure 45: Calibration Results, Standard Gas Tracer and GAS Menu

Item in Edit Mode Set to ALL

50
10.0
5

%LEL

%

ppm

CH

4

O

2

CAL

CO

800


ppm

CH

4

CAL

P
P
F

%LEL

%

ppm

CH

4

O

2

CAL

CO

F


ppm

CH

4

CAL

EDIT
ENTER