83 • Calibrating a Gas Tracer

Figure 44: Failed Fresh Air Adjustment, %Volume Sensor Installed and GAS Menu

Item in Edit Mode Set to ALL

NOTE:

If the GAS menu item in Edit Mode is set to Std, there will be no fresh
air adjustment result for the ppm combustible sensor and pressing and
releasing the EDIT ENTER button will have no effect.

In this case continue with step 28.

22. The SDM-2012 will apply GAS 1 calibration gas to the standard three sensors

for the time defined by the GAS TIME calibration parameter.