Setting the Operational Parameters in Edit Mode • 26

Setting the Gas Inlet for H

2

S (GX-2012 Only)

If you are using a separate calibration cylinder for the H

2

S channel, you may set

which gas inlet that cylinder will use. RKI Instruments, Inc. recommends using a 4-
gas mix cylinder that contains H

2

S and setting this parameter to GAS 1 (factory

setting).

To change the gas inlet for the H

2

S channel:

1.

Turn on the SDM-2012 with a GX-2012 and establish a connection between
them as described above in "Turning on the SDM-2012 with an Instrument".
The GX-2012 will display the home screen.

2.

Press and hold the EDIT ENTER button for 2 seconds. The GX-2012 will
display the Edit Mode screen.

3.

Press and release the EDIT ENTER button. The GX-2012 will alternate
between the following screens that show the current gas inlet for H

2

S along

with the current tested sensors setting. For instructions to change the tested
sensors setting, see “Updating the Tested Sensors Setting” on page 28.

50
12.0
50
25.0

%LEL

%

ppm

ppm

CH

4

O

2

TRANSMIT

CO

H

2

S




B / C / E

GAS 1
Std

EDIT



H2S
GAS

EDIT