Zero ng, Cal g2, Cal air – RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 86

Advertising
background image

81 • Calibrating a Gas Tracer

If the ppm sensor fails the fresh air adjustment, the SDM-2012 will abort
the calibration and will not apply calibration gas. If this happens, the
CAL

S

LED will flash red indicating a failure and the following screens will

alternate.

In this case continue with step 28.

19. The SDM-2012 will apply GAS 2 calibration gas to the ppm sensor for the time

defined by the GAS TIME calibration parameter. Be sure that the 1000 ppm
methane calibration cylinder is connected to the GAS 2 fitting.

20. The SDM-2012 will then apply fresh air to the standard three sensors and

%volume sensor (if installed and if the GAS menu item in Edit Mode is set to
ALL) for the time defined by the AIR TIME calibration parameter.

21. The SDM-2012 will perform a fresh air adjustment on the standard three

sensors and %volume sensor (if installed and if the GAS menu item in Edit
Mode is set to ALL).

If the air adjustment is successful, the SDM-2012 will continue as
described in step 22 below.