Calibrating a GX-2012 • 56

second and then release them. To avoid accidentally entering Edit Mode,
press and hold the POWER button first, then press and hold the EDIT
ENTER button.

CAUTION:

When using the GX-2012 with the SDM-2012, do not turn off the
instrument using the instrument power button. Use the EDIT ENTER
and POWER buttons on the SDM-2012 to turn off the instrument.

The CAL

S

LED will remain on indicating the test results. If the same GX-

2012 is turned on again, the test results will still be indicated by the CAL

S

LED and on the GX-2012 screen. To clear the CAL

S

LED, with the GX-

2012 on, press and release the EDIT ENTER button to return to the home
screen. If a new GX-2012 is turned on and connected, the results displayed
by the CAL

S

LED will automatically be cleared.

20. The results of the calibration will be stored in the SDM-2012’s memory and will

be available to copy to a USB flash drive. See “Copying Calibration and Bump
Test Records to a USB Flash Drive” on page 92 for instructions to copy the
saved bump test and calibration records to a USB flash drive.

21. Disconnect the gas out to instrument line and the exhaust line from the GX-

2012.

22. Remove the GX-2012 from the SDM-2012.

23. If you wish to calibrate additional instruments, repeat step 5 - step 22 above

for each additional instrument.