RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 38

Advertising
background image

33 • Connecting Calibration Gas

To connect calibration gas to the SDM-2012, do the following:

1.

If the area around the docking station is not considered a fresh air area (an
area free of combustible and toxic gases and of normal oxygen content,
20.9%) install a tube not longer than 10 feet on the filter attached to the AIR
fitting on the back of the docking station and route it to a fresh air area or
connect a cylinder of zero air with a demand flow regulator to the AIR fitting.

NOTE:

If you will be using a zero air cylinder with a Gas Tracer in the docking
station, you must connect the zero air cylinder to the AIR fitting using a
6 inch humidifier.

2.

Install the demand flow regulator on the calibration gas cylinder(s).

3.

Connect the demand flow regulator to the GAS 1 or GAS 2 inlet fitting using
the 3 foot length of 3/16 inch ID sample tubing provided with the docking
station.

For a 3- or 4-gas mix, connect the regulator to the GAS 1 inlet.

For a special calibration cylinder (ppm or %volume), connect the regulator to
the GAS 2 inlet keeping in mind that you will need to use a humidifier if bump
testing or calibrating a Gas Tracer.

ppm CH

4

/LEL CH

4

/Oxy/CO

(Gas Tracer)

• 3-gas mix with CH

4

/Oxy/CO

• 1000 ppm CH

4

ppm CH

4

/LEL CH

4

/%vol CH

4

/Oxy/CO

(Gas Tracer)

• 3-gas mix with CH

4

/Oxy/CO

• 1000 ppm CH

4

• 100% volume CH

4

Table 4: Recommended Gas Cylinders

Typical Instrument Types

Recommended Calibration Gas

Cylinder(s)