Fail c--limit cal c--limit – RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 43

Advertising
background image

Bump Testing a GX-2012 • 38

When CL RMNDR is set to ON and CL EXPRD is set to NOT USE, if the
GX-2012 is due for calibration, the instrument will show a calibration
expired failure screen and the buzzer and LED arrays will pulse.

The GX-2012 will then connect to the SDM-2012 and display the home
screen shown above in step 8. The CHARGE LED will begin to blink green.

When CL RMNDR is set to ON and CL EXPRD is set to CONFIRM, if the
GX-2012 is due for calibration, the instrument will show the following
screen and beep and flash the LED arrays for a few seconds indicating that
calibration is due.

When the beeping and flashing stops, press and release the RESET
SILENCE button on the GX-2012 to continue. The instrument will now
connect to the program, the home screen shown in step 8 above will be
displayed, and the CHARGE LED will begin to blink green.

10. If necessary, confirm that the bump test check tolerance is set to the desired

value. See “Setting the Bump Test Parameters” on page 21.

11. Verify that the appropriate calibration gas cylinder is connected to the GAS 1

and GAS 2 fittings on the back of the SDM-2012. See “Connecting Calibration
Gas” on page 32 for calibration gas cylinder options and calibration gas
connection procedures.

FAIL


C--LIMIT

CAL


C--LIMIT