P p f p, Edit enter – RKI Instruments SDM-2012 Standalone Configuration User Manual

Page 60

Advertising
background image

55 • Calibrating a GX-2012

Figure 30: Calibration Results, %Volume Sensor Installed and GAS Menu Item in

Edit Mode Set to ALL

NOTE:

If the GAS menu item in Edit Mode is set to Std, there will be no
calibration results for the %volume sensor and pressing and releasing
the EDIT ENTER button will have no effect.

19. After a successful or failed calibration,

To perform any other operations:

To perform a bump test, press and hold the BUMP

T

button until the pump

starts. To perform another calibration, press and hold the CAL

S

button

until the pump starts. To return to the home screen, press and hold the
EDIT ENTER button until the home screen appears.

To turn the GX-2012 off:

If the calibration was successful, the instrument will shut off after 15
seconds. If the calibration failed, the instrument will shut off after 10
minutes. If buttons are pressed before the SDM-2012 turns off the
instrument, it will automatically turn it off 10 minutes after the last button
push.

To turn off the instrument before it is automatically turned off, press and
hold the EDIT ENTER and POWER buttons simultaneously for about one