63 • Bump Testing a Gas Tracer

If the ppm sensor fails the fresh air adjustment, the SDM-2012 will abort
the bump test and will not apply calibration gas. If this happens, the
BUMP

T

LED will flash red indicating a failure and the following screens

will alternate.

In this case continue with step 31.

19. The SDM-2012 will apply GAS 2 calibration gas to the ppm sensor for the time

defined by the GAS TIME bump test parameter. Be sure that the 1000 ppm
methane calibration cylinder is connected to the GAS 2 fitting.

20. The SDM-2012 will analyze the bump test results.

If the ppm sensor passed the bump test, continue with step 21.

If the ppm sensor did not pass the bump test, and if AUTO CAL is set to
OFF, continue with step 21.

If the ppm sensor did not pass the bump test, and if AUTO CAL is set to
ON, a calibration will automatically begin and GAS 2 calibration gas will
continue to be applied. The CAL

S

LED will begin to flash amber.

2500


CH

4

ZERO NG

ppm

F


CH

4

ZERO NG

ppm

9800


ppm

CH

4

BUMP G2