Metrohm tiamo 1.3 Manual User Manual

Page 141

Advertising
background image

Subwindow Run

115

signed in a command then the window

Device assignment

appears for as-

signing the device.

•

Check solutions
Checks whether the solutions demanded in the device-specific commands
are present and available.

•

Check sensors
Checks whether the sensors demanded in the device-specific commands

are present and available.

Note
The Start test can also be triggered manually in the

READY

status with the menu

item

Tools, Run test

.

4 - Preconditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which
the option

Automatic conditioning

is switched on in the

START

command then pre-

conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (

Sample number

,

Statis-

tics

).

•

Conditioning requirement not fulfilled
After the start of preconditioning the status changes to

COND BUSY

. In this

status conditioning can be started with

[Start]

even though the conditioning

requirement has not been fulfilled. Conditioning can also be canceled with

[Stop]

or interrupted with

[Hold]

and then continued with

[Cont]

. If an error

occurs during conditioning then neither the Error track nor the Exit track will
be carried out.

•

Conditioning requirement fulfilled
If the conditioning requirement has been fulfilled then the status changes to

COND READY

. In this status the determination can be started with

[Start]

.

Conditioning can also be canceled with

[Stop]

or interrupted with

[Hold]

and

then continued with

[Cont]

. The conditioning requirement will continue to be

monitored continuously until the determination has been started.

5 - Start method

At the start of the method the following checks and actions are triggered:

•

Check sample data
Checks that the entered Sample data corresponds to that defined in the
Method variables and is valid.

•

Assign determination ID
An unambiguous determination ID is assigned for the determination.

•

Reserve device (module)
The devices (or their functional units) used by the device-dependent com-
mands contained in the method are reserved for the duration of the deter-

mination, i.e. they cannot be used in other workplaces nor can they be op-
erated manually.

•

Increment sample number
The

Sample number

is increased by

+1

.

•

Increment statistics counter
If statistics has been switched on in both the

START

command and on the

tab

Single determinations

then the statistics counter is increased by

+1

. If a

new method is loaded or if the statistics counter agrees with the set counter
then the current statistical data will first be deleted and the statistics counter
reset to

0

.