2 login/password protection, 1 general – Metrohm tiamo 1.3 Manual User Manual

Page 36

Advertising
background image

10

General program functions

2.2 Login/Password

protection

2.2.1 General

Login in tiamo

tiamo can be configured so that all users must log in with User name and Pass-
word
; these entries are then checked automatically. A requirement for this is that
a User administration has been set up and the corresponding Security settings

have been made. This data is stored in the configuration database. In client/server
systems this is found on the server and applies globally for all clients (central user
administration).

FDA-conform settings

If FDA-conform work is to be carried out then the settings on the tab
Login/Password protection in the dialog window

Security settings

as per 21 CFR

11 must be switched on with the button

[Set]

. The following conditions are then

observed:

•

Each time that the program is started a Login with user name and

password is required.

•

Password administration takes place in tiamo.

•

User names must be unambiguous. Once users have been entered they
cannot be deleted.

•

Passwords must be unambiguous for each user. No password that has
been used once by the user and has expired may be reused.

•

Passwords must have a minimum number of characters.

•

Passwords must have a defined Validity period after which they must be
changed.

•

The number of incorrect attempts for entering the password is limited. If
this number is exceeded then the user is automatically given the status

in-

active

.

Actions

If login is switched on then the following actions can be carried out:

•

Login at program start

•

Manual logout

•

Automatic logout

•

Change password