Subwindow Devices

765

RS 232

COM Port

COM1, COM2, ..., [ first free COM port ]

Selection of the serial interface on the PC to which the device is connected.

Baud rate

300, 600, 1200, 2400, 4800, [ 9600 ], 19200, 38400, 57600, 115200

Transmission speed.

Data bit

7, [ 8 ]

Number of data bits.

Parity

[ None ], Odd, Even

Type of parity check.

Stop bit

[ 1 ], 2

Number of stop bits.

Handshake

[ None ], HW, SW

Type of data transmission protocol.

Timeout

1000 ... [ 2000 ] ... 9000 ms

Maximum waiting time for receiving characters. If this time is exceeded the
reception will be stopped.

Terminator for send

20 characters, \0D, \0D\0A, \0D\0A\0A, \0D\0D\0A

(

\0D

= CR,

\0A

= LF)

Final character for transmission in hexadecimal code.

Terminator for receive

20 characters, \0D, \0D\0A, \0D\0A\0A, \0D\0D\0A

(

\0D

= CR,

\0A

= LF)

Final character for reception in hexadecimal code.

Code page

[ Cp437 ], Cp850, Cp852, ASCII

Code page used for the data transmission.

Send characters one by one

on, [ off ]

If this option is enabled every character is sent separately. Before sending

the next character, the

Delay time

is waited for.

Delay time

[ 0 ] ... 9000 ms

Waiting time before sending the next character.