Commands

457

Action

Definition of the action to be taken when the limits of the temperature are in-

fringed:

Stop determination

(only Titrando)

The running

STAT

command is quit, then the exit track (if present) is started

and the determination is stopped.

Stop command

The running

STAT

command is quit, then the next command is started.

Wait for [Continue]

Reagent dosing in the current

STAT

command is interrupted and a message

appears. As soon as the monitored temperature is again within the limits
(including hysteresis) reagent dosing can be resumed by pressing the

[Continue]

button in this message window.

Wait for limit ok

Reagent dosing in the current

STAT

command is interrupted. As soon as the

monitored temperature is again within the limits (including hysteresis) re-
agent dosing is resumed automatically.

[ none ]

No action will be taken.

Track call on limit exceeding

In this table, which cannot be edited directly, max. 20 entries can be de-
fined for the track that is to be started automatically when a particular limit
is infringed.

Open the dialog window

Track call #

(see below) to enter parameters of a

new track call.

Open the dialog window

Track call #

(see below) to edit parameters of the

selected track call.

Delete the selected track call.

Track call

Monitoring

[ Measured value ], Dosing rate, Temperature, Any

Selects the quantity for which a track is to be started if its limits are infringed

(

Any

= any of the three quantities).

Limit exceeding

Lower limit, Upper limit, [ Any ], OK

Selects the limit infringement for which a track is to be started. With

Any

the

track will be started for an infringement of either the lower or the upper limit,