Metrohm tiamo 1.3 Manual User Manual

Page 560

Advertising
background image

534

Method

The calibration can be carried out by manual changing of buffers/standards or by
automatic changing in combination with a Sample Processor.

Commands

The following commands can be selected for calibration:

•

CAL LOOP pH

Calibration loop for calibrating pH electrodes.

•

CAL MEAS pH

Measuring command for calibrating pH electrodes.

•

CAL LOOP Conc

Calibration loop for calibrating ion selective electrodes (ISE electrodes).

•

CAL MEAS Conc

Measuring command for calibrating ion selective electrodes (ISE elec-
trodes).