Metrohm tiamo 1.3 Manual User Manual
Page 418

392
Method
SET
Command for Set Endpoint Titrations (
SET
).
Principle
With this command titrations to one or two preset endpoints are carried out. The
termination of the titration at the endpoint is either drift-controlled or after a waiting
period. The volume that has been added when the endpoint is reached gives the
calculable reagent consumption.
Reagent addition takes place in three phases during the titration:
•
Initial addition
During this phase the dosing rate increases continuously. It starts with the
Minimum rate
and increases to the
Maximum rate
.
•
Continuous addition
In this phase addition is carried out at the
Maximum rate
until the
Control
range
is reached.
•
Control range
In this range addition is finely controlled. Shortly before the end point is
reached addition is only carried out at the
Minimum rate
(see "Control
range").