564

Method

Filling rate

0.01 ... 166.00 mL/min, [ maximum ]

(only Titrando, 855)

Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the ex-
change unit or dosing unit used. If the entered filling rate is too high for the
selected dosing device then during dosing it will automatically be reduced
to the largest possible value (see STAT pH - General/Hardware).

Sensor

Measuring input

[ 1 ] ... 2

(835, 836, 842, 857)

[ 1 ]

(855)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ Metal electrode ]

Selection of a sensor of type

pH electrode, Metal electrode

or

ISE electrode

from the list of sensors defined in the sensor table. The calibration data for
the ISE and pH sensors will be adopted for the determination.

Temperature measurement

Type of temperature measuring (only Titrando):

continuous

A temperature sensor must be connected. The temperature is measured
continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered

Temperature

(tab

Titra-

tion parameters

) will used.

off

The temperature will not be measured. The

Temperature

(see Dosing pa-

rameters) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(only

Titrando)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(only

Titrando)

The stirrer is automatically switched off at the end of the command if this
option is enabled.

Dosing parameters

Screenshot and parameters: see DOS pH - Dosing parameters

Stop conditions

Screenshot and parameters: see DOS pH - Stop conditions