Metrohm tiamo 1.3 Manual User Manual
Page 558

532
Method
opened with the menu item
Determinations, Show calibration curve...
or the symbol
.
General/Hardware
Screenshot and parameters: see STDADD dos - General/Hardware
Standard addition
The standard addition parameters are defined on this tab.
Command name
25 characters, [ STDADD auto # ]
Name of the command.
Number of additions
1 ... [ 3 ] ... 19
Number of times the standard solution is to be added.
Volume auxiliary solution
[ 0.000 ] ... 9999.999 mL
Volume of auxiliary solution (e.g. buffer) that is to be added to the sample
before the first standard addition.
Stop volume
0.0 ... [ 100.0 ] ... 9999.9 mL, off
Stops when the sum of the added volume increments exceeds the stop
volume defined here.
Addition solution
(only displayed for
Solution
=
not defined
)
Concentration
0.001 ...[ 1.000 ] ... 999999.999
Concentration of addition solution.
Unit