Metrohm tiamo 1.3 Manual User Manual

Page 488

Advertising
background image

462

Method

Threshold value

0.1 ... [ 1.0 ] ... 20.0 pH/mL

The evaluation begins as soon as the slope of the curve exceeds the set

threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 pH/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Additional measured values

Abbildung und Parameter: siehe DET pH - Weitere Messwerte

STAT U

Command for STAT titrations with measured value U.

Devices

This command can be executed with the following devices which have the

STAT

pH

mode:

Titrino: 718, 736, 751, 799

Titrando: 835, 836, 842, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

STAT U

are configured in the following 8 tabs:

•

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.

•

Start conditions

Parameters for the measurement of the initial measured value and for defin-
ing the conditions which have to be met at the start of the titration.

•

Control parameters

Parameters for the control point.

•

Titration parameters

Parameters for the run of the titration.

•

Stop conditions

Definitions of conditions which cause the stop of the titration.

•

Monitoring

Definition of measured value, dosing rate, and temperature monitoring.

Advertising