Metrohm tiamo 1.3 Manual User Manual
Page 551

Commands
525
25 characters, [ STDADD man # ]
Name of the command.
Number of additions
1 ... [ 3 ] ... 19
Number of times the standard solution is to be added.
Volume auxiliary solution
[ 0.000 ] ... 9999.999 mL
Volume of the auxiliary solution (e.g. buffer) to be added to the sample be-
fore the first standard addition.
Addition mode
Fixed volume increments
[ on ], off
If this option is enabled a fixed volume increment muss be added for each
standard addition.
Addition volume
[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL
Quantity of the fixed volume increment.
Variable volume increments
on, [ off ]
If this option is enabled the defined volume increment muss be added for
each standard addition.
Addition volume 1...19
[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL
Quantity of the defined volume increments.
Addition solution
Concentration
0.001 ...[ 1.000 ] ... 999999.999
Concentration of addition solution.
Unit
mol/L, mmol/L, µmol/L, [ ppm ], g/L, mg/L, µg/L, mg/mL, µg/mL, %, mEq/L
Concentration unit of addition solution.
Measuring parameters
The measuring parameters are defined on this tab.