Bio-Rad Microplate Manager Software User Manual
Page 96
![background image](/manuals/600349/96/background.png)
Microplate Manager User Guide
86
Kinetic Correlation Coefficient Report
Figure 60. Kinetic Correlation Coefficient Report.
The Kinetic Correlation Coefficient Report displays the correlation coefficient of
the velocity segment for each well. The correlation coefficient, or “r” value, is a
measure of the precision of the fit of the linear regression to the raw data. An “r”
value of 1.00 represents a perfect correlation. See the discussion in section 6.8.
Kinetic Standard Errors Report
Figure 61. Kinetic Standard Errors Report.
The Kinetic Standard Errors Report shows the calculated standard error of the
velocity for each well. If you have large standard errors (more than 5–10% of the
corresponding velocity), you should reset the limits (under OPTIONS > Analysis
Options...) to focus on the linear portion of the results.