Error message list – KEYENCE LJ-H1W User Manual
Page 93
Error Message List
Appendix-9
LJ-H1W-M-NOA-E
A
Error Message List
This section explains information about the displayed error messages that appear when an error occurs.
Item
Error message
Measure
Reference
File operations Failed to read the file ( ).
Check that the correct file is set.
* The path name is entered in the
parentheses ( ).
Check whether the selected file is an LJ-
Navigator program file (*.cfg) or an
environment setting file (*.env).
Page 3-2
Failed to read the environment settings.
( )
* The settings that could not be read are
entered in the parentheses ( ).
Check whether the unreadable environment
setting items are set correctly on the
controller.
Page 3-4
Sending and
receiving
settings
Failed to send settings.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Page 3-6
Failed to receive settings.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Page 3-6
Head setting
Failed to acquire profile.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Page 4-7
Failed to start calibration.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
Page 5-3
Failed to pause calibration.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
• Set the measurement target correctly
within the measurement range and
restart calibration.
_
Failed to stop calibration.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
_
Failed to initialize calibration.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement
screen (Measuring)".
_
Set within valid range.
An incorrect value is set in the window of a tab.
Set the value within the valid range.
Page 5-3
The value for sampling count exceeds the
setting range.
Set the value between 1 and 100.
Set the sampling count between 1 and 100.
Page 5-3