24

BM

series

6.4.

Percent Mode (%)

The percent mode displays the weight value in percentage compared to a 100% reference mass

and is used for target weighing or checking the sample variance.

Selecting the percent mode

1 Press

the

MODE

key to select the unit

%

(Percent mode).

If the percent mode can not be selected, refer to "5. Weighing Units".

Storing the 100% reference mass

2 Press

the

RANGE

key to enter the 100% reference mass

storing mode.

3 Place a tare (container) on the weighing pan, if necessary.

Press the

RE-ZERO

key to cancel the weight (tare).

The balance displays

100 0 %

.

4 Place the sample to be set as the 100% reference mass on

the pan or in the container.

5 Press

the

PRINT

key to store the reference mass.

The balance displays

100.00 %

.

Note

The decimal point position can be changed by the 100% mass.

BM-252, BM-500

BM-300, BM-200

BM-20, BM-22

100% mass

Minimum

display

100% mass

Minimum

display

0.0100 g to 0.0999 g 1 %

0.00100 g to 0.0099 g 1 %

0.1000 g to 0.9999 g 0.1 %

0.01000 g to 0.0999 g 0.1 %

1.0000 g to

0.01 % 0.10000 g to

0.01 %

If the balance judges that the mass of the sample is too light

(under 0.01g) to be used as a reference, it displays

lo

.

A 100% reference mass can be stored in the non-volatile

memory and is maintained even if the AC adapter is removed.

6 Remove

the sample.


Reading the percentage

7 Place a sample to be compared to the reference mass on

the pan. The displayed percentage is based on the 100%
reference mass.

100% mass

Place sample

Percentage

Weighing pan

Tare

(Container)