A&D Weighing BM-20 User Manual

Page 68

Advertising
background image

66

BM

series

Measuring the density

(

specific gravity

)

Note If the liquid temperature or the type of liquid is changed during measurement, again

input the value of the liquid density as necessary. The density display is four decimal
places. The minimum display can not be changed with the

RANGE

key.


The density (or specific gravity) is displayed after "Mass

measurement in air" and "Mass measurement in liquid".
The procedure of each measurements is as follows:
g : gram. : the processing indicator.

1 Enter the density mode. (g lights. blinks.)

Place nothing on both pan and press the

RE-ZERO

key

to display zero.


2 Place the sample on the pan in air. (g lights. blinks.)

If the weight value is stored or output, press the

PRINT

key to store it after a stable weight value is displayed.
Press the

RANGE

key to decide the weight value

in air and proceed to next step.

Note If negative value or e (out of range) is displayed,

the

RANGE

key is inactive.




3 Move the sample to the pan in liquid. (g lights. lights.)

If the weight value is stored or output, press the

PRINT

key to store it after a stable weight value is displayed.
Press the

RANGE

key to decide the weight value in

liquid and proceed to next step.

Note If e (out of range) is displayed, the

RANGE

key

is inactive.



4 If the density value is stored or output, press the

PRINT

key to store it. (Unit : .

g turns off. lights.)

Press the

RANGE

key to measure another sample

and proceed to step 2.



5 If the liquid temperature or the type of liquid is changed

during measurement, input again the value of the liquid
density as necessary.



6 Press

the

MODE

key to proceed to other modes.

From "Entering the

density of a liquid"

Place sample on

the pan in air

Move sample from

the pan in air to the

pan in liquid

Remove sample

Output or store

Output or store

Output or store

Weighing in air.

g lights. blinks.

Weighing in air.

g lights. blinks.

Weighing in liquid.

g lights. lights.

Density display.

g turns off. lights.

Advertising
This manual is related to the following products: