Calibration using an external weight – A&D Weighing BM-20 User Manual

Page 36

Advertising
background image

34

BM

series

8.5.

Calibration Using An External Weight

This function calibrates the balance using an external weight.


1 Connect the AC adapter and warm up the balance for at

least one hour with nothing on the weighing pan.


2 Press and hold the

CAL

key until

Calout

is

displayed, then release the key.


3 When

displaying

Cal 0

, proceed as follows:

If you want to change the calibration mass, press

the

RANGE

key and proceed to step 4.

If you use the calibration mass value stored in

the balance, proceed to step 5.


4 Specify the calibration mass value as follows:

RANGE

key..........The key to switch blinking figures.

RE-ZERO

(+)key.....The keys to select the calibration mass or

adjust the mass value. Refer to page 31.

PRINT

key............The key to store the new mass value.

Even if the AC adapter is removed, the
data is maintained in non-volatile memory.

CAL

key ...............The key to cancel the operation and

return to

Cal 0

.


5 Confirm that there is nothing on the pan and press the

PRINT

key. The balance measures the zero point.

Do not allow vibration or drafts to affect the balance.

The balance displays the calibration mass value.


6 Place the displayed calibration weight on the pan and press

the

PRINT

key. The balance measures the calibration mass.

Do not allow vibration or drafts to affect the balance.


7

end

is displayed.

Remove the weight from the pan.


8 If GLP output is set,

glp

is displayed and the calibration

report is output and stored. Refer to "GLP output (

info

)" and

"Data memory (

data

)" of the function table, "

11.2. GLP

Report" and "12. Data Memory".


9 The balance will automatically return to the weighing mode.

10 Place the calibration weight on the pan and confirm that the

value displayed is correct. If it is not within the range, check
the ambient conditions such as breeze and vibration also
check the weighing pan. Then, repeat steps 1 to 10.

MODE

(-)key

Press and hold

Release

Select a type

Place the weight

Remove the weight

Specify detail of value

GLP output

Example:

New value

100.0012 g

New mass

value