16. Maintenance

16.1.

Treatment Of The Balance

Clean the balance with a lint free cloth that is moistened with warm water and a mild detergent.
Do not use organic solvents to clean the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.
Use the original packing material for transportation.
Consider

"

3. Precautions" when operation the balance.

16.2.

Error Codes

Display Error

code

Description

EC,E00

Internal error
If displaying this error continuously, contact your local A&D dealer to
repair the balance.

EC,E11

Stability error
The balance can not stabilize due to an environmental problem. Check
around the pan. Prevent vibration, drafts, temperature changes, static
electricity and magnetic fields, from influencing the balance. Press the

1

CAL

1

key to return to the weighing mode.

EC,E16

Internal mass error
Applying the internal mass does not yield a change in the mass value as
specified. Confirm that there is nothing on the pan and perform the
weighing operation from the beginning again.

EC,E17

Internal mass error
The internal mass application mechanism does not function properly.
Perform the weighing operation from the beginning again.

EC,E20

Calibration weight error
The calibration weight is too heavy. Confirm the calibration mass value.
Press the

1

CAL

1

key to return to the weighing mode.

EC,E21

Calibration weight error
The calibration weight is too light. Confirm the calibration mass value.
Press the

1

CAL

1

key to return to the weighing mode.

Overload error
A sample beyond the balance weighing capacity has been placed on the
pan. Remove the sample from the pan.

Weighing pan Error
The mass value is too light.
Confirm that the weighing pan is properly installed and calibrate the balance.

Sample mass error
The balance can not store the sample for the counting mode or for the
percent mode because it is too light. Use a larger sample.