6.7.

Repeatability Test

The repeatability means the weighing error when measuring the same mass repeatedly.

The standard deviation is generally used for the indicator of the repeatability.

The repeatability test measures the built -in mass 10 times and displays the standard deviation.
Use the repeatability test to inspect the performance of the balance when changing environment

or when moving the balance to a new location.

Example: "The standard deviation = 0.2

mg" means that the probability of weighing error within

±0.2

mg is approx. 68% when measuring the same mass repeatedly.

Note

The repeatability test uses the internal mass. Therefore, the standard deviation discords to item

repeatability of "

17.Specifications". Regard the standard deviation as a reference.
Products Internal

mass

BM-20, BM-22

Approx.

20

g

BM-252, BM-500, BM-300, BM-200

Approx.

190

g



1 Press and hold the

RANGE

key,

Release the key when displaying

rep te5t

.


2 When

displaying

rep te5t

, the measurement

is automatically started,

"r"e"p

blinks and the

measurement count (0 to 10) is displayed.
When stopping the current measurement,
press the

CAL

key.

Note

Avoid vibration and breezes for precise

measurement during this test.


3 At the end of the measurement,

the standard deviation of the
repeatability is displayed.


4 Press

the

CAL

key or

PRINT

key

to return to weighing mode.


Press and hold

Release

In measurement

Stop

or

Result

The standard deviation