Glp report, Calibration report using the internal mass – A&D Weighing BM-20 User Manual

Page 55

Advertising
background image

BM series

53

11.2.

GLP Report

Set the following parameters to output the report.

Refer

to

"

15.2. Connection to peripheral equipment" for connection to an AD-8121B.

To print the report, set the "GLP output (

info

)" parameter to "

1

" and use MODE 3 of the AD-8121B. If

MODE1 is used, select temporary dump print mode by pressing the

STAT.

key of the AD-8121B.

To output the report to a personal computer using the RS-232C interface, set the "GLP output (

info

)"

parameter to "

2

".

If the time and date are not correct set the correct time and date in "Clock (

Cl adj

)" of the function table.



Calibration report using the internal mass

Key operation
1 Press

the

CAL

key to display

Cal in

and calibrate the balance automatically.

2 If GLP output is used,

glp

is displayed and the calibration report is output.

3 The balance returns to weighing mode after this calibration.

Setting

of

"info 1" Setting

of

"info 2"

AD-8121 printer format General

format














Space, ASCII 20h

<TERM>

Terminator, CR , LF or CR

CR

Carriage

return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration type

Signature

===========

A

=

&

=

D

MODEL

=====

BM-300

S/N

=====

01234567

ID

======

LAB-0123

DATE

==

2011/02/20

TIME

====

12:34:56

CALIBRATED(INT.)
SIGNATURE


-

-

-

-

-

-

-

-

-

-

-

-

-

-

-