Details of the function table – A&D Weighing BM-20 User Manual

Page 43

Advertising
background image

BM series

41

10.2.

Details Of The Function Table

Class Item

and

Parameter

Description

0 Fast response, sensitive value

FAST

1

MID.

Cond
Condition

2 Slow response, stable value

SLOW

Can be changed by

response adjustment.

0 Stable range is ±1 digit

1 ±2

digit

#1

5t-b
Stability band width

2 Stable range is ±3 digits

The stabilization indicator

illuminates when the display

fluctuation is within the range.

off OFF

ploc
Display lock function on ON

Display hold function to measure the

volume dispensed from the pipette.

0 OFF

1 Normal

2 Strong

trc
Zero tracking

3 Very strong

Keeps zero display by tracking

zero drift.

0 5 times/second

5pd
Display refresh rate 1 10 times/second

Period to refresh the display

0 Point (.)

pnt
Decimal point

1 Comma (,)

Decimal point format

off OFF

p-on
Auto display-ON on ON

Turns on the weighing mode display

when the AC adapter is connected.

off OFF

ba5fnc
Environment
Display

beep
Beep

on ON

Turns on the weighing mode display

when the AC adapter is connected.

Cl adj

Clock

Refer to "

10.7. Clock and Calendar Function"

The time and date are added to

the output data.

0 Key mode

Accepts the

PRINT

key only

when the display is stable.

1 Auto print mode A

Reference = zero

2 Auto print mode B

Reference = last stable value

Outputs data when the display is

stable and conditions of ap-p,

ap-b and the reference value are

met.

prt
Data output mode

3 Stream mode /

Interval mode

With data 0, outputs data

continuously; with data 2, uses

interval memory.

0 Plus only

Displayed value > Reference

1 Minus only

Displayed value < Reference

ap-p
Auto print polarity

2 Both

Regardless of displayed value

0 10 digits

1 100 digits

ap-b
Auto print

difference

2 1000 digits

Difference between reference

value and displayed value

#2

0 Not used

1 Stores weighing data

dout
Data output

data
Data memory

2 Stores calibration data

Related items:

prt, int, d-no, 5-td, info

: Factory settings.

#1

The unit of minimum display is digit.

Example: If 1 mg display is selected using the

RANGE

key for the

BM-300, 1 mg is one digit.

#2

Usable minimum display of the balance is one digit.

Example: In gram display, one digit is 0.01 mg for the

BM-252 and 0.1 mg for the BM-300.