Calibration, Calibration group – A&D Weighing BM-20 User Manual

Page 33

Advertising
background image

BM series

31

8. Calibration

(To Adjust The Weighing Value)

8.1.

Calibration Group

Calibration

Automatic

self

calibration (Calibration due to changes in temperature)

Calibration

using

the internal mass (One-touch calibration)

Calibration

using

an external weight that you have

Calibration test

Calibration

test

using the internal mass

Calibration test using target mass that you have

Correction of the internal mass value

Correction of the internal mass value

Caution

Do not allow vibration or drafts to affect the balance during calibration.
Calibration test does not perform calibration.
When using the data output for GLP using the RS-232C interface, set "GLP output (

info

)" of the

function table. Refer to "

10. Function Table". Time and date can be added to the GLP report. If it

is incorrect, refer to "

10.7 Clock and Calendar Function" and adjust them.

Calibration

test

is available only when "GLP output (

info

)" is set .

The calibration and calibration test data can be stored in memory. When using memory, set "Data

memory (

data

)" of the function table. Refer to "

12. Data Memory" for details.

Caution on using an external weight

The accuracy of an external weight can influence the accuracy of weighing.
Select a mass for calibration and calibration test from the following table.

Model Usable

calibration

mass Adjustable

range

BM-20

1 g

2 g

5 g

10 g

20 g*

BM-22

1 g

2 g

5 g

10 g

20 g*

-3.000 mg to +3.099 mg

BM-252

10 g

20 g

50 g

100 g

200 g*

-15.00 mg to +15.99 mg

BM-200

50 g 100 g

200 g*

BM-300

50 g 100 g

200 g* 300

g

BM-500

50 g 100 g

200 g*

300 g

500 g

-30.0 mg

to +30.9 mg

Bold type*: Factory

setting.

The calibration mass value can be adjusted within the range above.

About the internal mass

The internal mass may change due to corrosion or other damage caused by the operating

environment, or due to aging. Check the internal mass periodically and correct the internal mass
value if necessary. Refer to "

8.7. Correcting the Internal Mass Value".

Display

This

indicator

means

"In process of measuring calibration data".

Do not allow vibration or drafts to affect the balance while the indicator is
displayed.