7. Response

Adjustment

This function stabilizes the weight value by reducing the influence on weighing that is caused

by drafts and/or vibration at the place where the balance is installed. The adjustment of the
function can be performed by analyzing the environment automatically or hand-operation.
The state of the function has three stages as follows :

Indicator Parameter Response

Stability

FAST

Cond 0

MID.

Cond 1

SLOW

Cond 2

Fast response, Sensitive value

Slow response, Stable value



7.1.

Automatic Response Adjustment

This function automatically updates the response adjustment by analyzing the influence of

the environment using the internal mass.


1 Press and hold the

MODE

key until

RESPONSE

is

displayed, then release the key.

2 The balance automatically sets the response characteristic.

Caution Do not allow vibration or drafts to affect the

balance during adjustment.

3 After automatic adjustment, the balance displays

end

,

returns to the weighing mode and displays the updated
response indicator for about thirty seconds.

Note

If the automatic response adjustment fails, the balance

displays

CH ng

. Check the ambient conditions such as

breeze and vibration, also check the weighing pan. Then,
perform the adjustment again. Press the

CAL

key to

return to the weighing mode.

If there is anything on the weighing pan, the balance

displays

CH 0

. Remove the substance from the pan.

Press the

CAL

key to return to the weighing mode.


Advise
If the automatic response adjustment is not helpful, try "

7.2.

Manual Response Adjustment".