Title block and end block – A&D Weighing BM-20 User Manual

Page 59

Advertising
background image

BM series

57

Title block and end block

When mass values are recorded as GLP data, a "Title block" is inserted at the beginning and an

"End block" is inserted at the end of a group of mass values in the GLP report.

Notes

To output the report to an AD-8121B , use MODE 3 of the AD-8121B. If MODE1 is used, select

temporary dump print mode by pressing the

STAT.

key of the AD-8121B.

If the data memory function is used (except

data 0

), the "Title block" and "End block" can not be

output.


Key operation
1 With

the

weighing data displayed, press and hold the

PRINT

key, then release the key.

5tart

is displayed. The "Title block" is output.

2 The weighing data is output according to the parameter setting of the data output mode (

prt

) of

the function table.

3 Press and hold the

PRINT

key until

recend

is displayed, then release the key.

The "End block" is output.


Setting

of

"info 1" Setting

of

"info 2"

AD-8121 printer format General

format

























Space, ASCII 20h

<TERM>

Terminator, CR , LF or CR

CR

Carriage

return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

-----------

A

-

&

-

D

MODEL

-----

BM-300

S/N

-----

01234567

ID

------

LAB-0123

DATE

--

2011/02/20

START
TIME

----

12:34:56



WT

--

+123.4567

--

g

WT

--

+123.4612

--

g

WT

--

+123.4623

--

g

WT

--

+123.4634

--

g

WT

--

+123.4511

--

g

WT

--

+123.4534

--

g

WT

--

+123.4712

--

g

WT

--

+123.4645

--

g


END

TIME

----

12:40

15
SIGNATURE


-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

Title block

Manufacturer

Model

Serial number

ID number

Date

Time





Weighing data






Time

Signature



End block