Other display – A&D Weighing BM-20 User Manual

Page 80

Advertising
background image

78

BM

series

Display Error

code

Description

EC,E02

Not ready
A received command can not be processed.
Example:

The balance received a "Q" command, but not in the weighing mode.
The balance received a "Q" command while processing a RE-ZERO

command. Adjust the delay time to transmit a command.

EC,E03

Timeout error
If the timeout parameter is set to "

t-Up on

", the balance did not receive

the next character of a command within the time limit of one second.
Confirm the communication.

EC,E04

Excess characters error
The balance received excessive characters in a command.
Confirm the command.

EC,E06

Format error
A command includes incorrect data.
Example:

The data is numerically incorrect. Confirm the command.

EC,E07

Parameter setting error
The received data exceeds the range that the balance can accept.
Confirm the parameter range of the command.

Other error code

If an error described above can not be released or other errors are
displayed, contact the local A&D dealer.



16.3.

Other Display

When this indicator blinks, automatic self calibration is required. The indicator
blinks when the balance detects changes in ambient temperature. If the balance
is not used for several minutes with this indicator blinking, the balance performs
automatic self calibration. The blinking period depends on the operating
environment.

Advise

The balance can be used while this indicator is blinking. We recommend that you
perform automatic self calibration for precision weighing.