Calibration Test Using An External Weight

This function tests the weighing accuracy using an external weight.
Calibration test report can be output or stored with "GLP output (

info

)"

Calibration test does not perform calibration.


1 Connect the AC adapter and warm up the balance for at least

one hour with nothing on the weighing pan.

2 Press and hold the

CAL

key until

CC out

is displayed,

then release the key.

3 When

displaying

CC 0

, proceed as follows:

If the target mass is changed, press the

RANGE

key and

proceed to step 4. A list of usable weights is on page 31.

If current target mass value is used, proceed to step 5.


4 Specify the target mass value as follows:

RANGE

key........ The key to switch blinking figures.

RE-ZERO

(+)key... The keys to select the target mass or

adjust the mass value. Refer to page 31.

PRINT

key.......... The key to store the new mass value.

Even if the AC adapter is removed, the
data is maintained in non-volatile memory.

CAL

key............. The key to cancel the operation and

return to

CC 0

.


5 Confirm that there is nothing on the pan and press the

PRINT

key. The balance measures the zero point and

displays the measured value. Do not allow vibration or
drafts to affect the balance. The balance displays the
target mass value.



6 Place the displayed target mass on the pan and press the

PRINT

key. The balance measures the target mass and

displays the measured value. Do not allow vibration or
drafts to affect the balance.



7

end

is displayed.

Remove the weight from the pan.


8 If GLP output is set,

glp

is displayed and the calibration

report is output and stored. Refer to "GLP output (

info

)" and

"Data memory (

data

)" of the function table, "

11.2. GLP

Report" and "12. Data Memory".


9 The balance will automatically return to the weighing mode.

MODE

(-)key

Press and hold

Release

Select a type

Place target mass

Remove target mass

Specify detail of value

GLP output

Example:

New value

100.0012 g

New mass

value