Specifications – A&D Weighing BM-20 User Manual

Page 82

Advertising
background image

80

BM

series

17. Specifications

BM-20 BM-22 BM-252

BM-500

BM-300

BM-200

22 g

Weighing capacity

22 g

5.1 g

250 g

520 g

320 g

220 g

22.00008 g

Maximum display

22.000084 g

5.100009 g

250.00084 g 520.0084 g 320.0084 g 220.0084 g

0.01 mg

Minimum display

0.001 mg

0.001 mg

0.01 mg

0.1 mg

0.01 mg

Repeatability

(Standard

deviation) 0.0025 mg

0.004 mg

0.03 mg

0.2 mg

0.1 mg

±0.02 mg

Linearity ±0.010

mg

±0.010 mg

±0.10 mg

±0.5 mg

±0.3 mg

±0.2 mg

Approx. 8 seconds

Stabilization time

(Typical

at

FAST

)

Approx. 10 seconds

Approx. 3.5 seconds

Sensitivity drift

±2 ppm/°C (10 °C to 30 °C, Automatic self calibration: OFF)

Operating environment

5 °C to 40 °C, 85 %RH or less (No condensation)

Display refresh rate

5 times/second or 10 times/second

Environmental sensors

Resolution: Temperature ±1.5 °C, Humidity ±10 %, Air pressure ±10 hPa

Applicable range: 5 °C to 40 °C

Minimum unit

mass

0.1 mg

Counting

mode Number of

samples

10, 25, 50 or 100 pieces

Minimum 100%

reference mass

1 mg

10 mg

Perc

ent

mode Minimum display

0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)

Interface USB

and

RS-232C with Windows Communication Tools Software WinCT

External calibration

weight

20 g

10 g

5 g

2 g

1 g

20 g

10 g

5 g

2 g

1 g

200 g

100 g

50 g

20 g

10 g

500 g

300 g

200 g

100 g

50 g

300 g

200 g

100 g

50 g

200 g

100 g

50 g

Size of loading area for

weighing mm

Weighing pan:

φ25,

Filter weighing:

φ50, φ95

Test tube holder:

φ12

Weighing

pan:

φ90

Test tube holder:

φ12 (Option)

External dimensions

259(W) x 466(D) x 326(H) mm

Power supply &

AC adapter type

Power consumption: Approx. 30VA (supplied to the AC adapter )

Confirm that the adapter type is correct for the local voltage and power

receptacle type.

Mass Approx.

10

kg

The

operating

environment does not include excessive change of ambient temperature, humidity,

vibration, drafts, magnetic fields and static electricity.

The

internal

mass

may change due to corrosion or other damage caused by the operating environment,

or due to aging. Check the internal mass periodically and correct the internal mass value if necessary.

∗1

When placing the tare weight within the standard range, if using the smart range, the weighing sample

can be weighed within the precision range.

∗2

The repeatability for 1 gram within the weighing capacity.

∗3

The repeatability for 100 gram within the weighing capacity.

∗3

∗1

∗2

∗2