Monitoring clock and environmental sensors – A&D Weighing BM-20 User Manual

Page 29

Advertising
background image

BM series

27

6.6.

Monitoring Clock and Environmental Sensors

In the weighing mode, the built-in clock and environmental sensors (temperature, humidity, air
pressure) can be monitored and temperature data can be output with the function settings.

Specifications Of The Environmental Sensors

Sensor

Resolution Measurement range

Applicable range

Temperature

±1.5 °C 5 °C to 40 °C

Humidity

±10 %

0 % to 100 %

Air pressure

±10 hPa

300 hPa to 1100 hPa

5

°C to 40

°C

Caution

The environmental sensors are built into the case. Therefore, these values are not in accordance

with the environmental condition.

Refer

to

"

17. Specifications" for operating condition of the product.

Operation

Press the

SELECT

key in the weighing

mode. The current time, date and sensor
values (temperature, humidity, air pressure)
are displayed for several seconds in order.
The balance returns to the weighing mode
automatically.

If pressing the

SELECT

key while

displaying current data, the display
proceeds to next data.

If pressing the

CAL

key while

displaying current data, the balance
returns to weighing mode.

Note

If adjusting the built-in clock, refer to "

10.7.Clock And Calendar Function".

If appending time and date to data output, refer to the [Time/Date output (5-td)] of

"

10.2.Details Of The Function Table".

If appending temperature data to the data output, refer to [Temperature output (5-tp)] of

"

10.2.Details Of The Function Table".


Humidity

Air pressure

Temperature

Weighing mode

Time

5 seconds

Date

5 seconds

Environmental sensors

10 seconds

Weighing mode