Entering the density of a liquid, Entering the water temperature, Entering the density directly – A&D Weighing BM-20 User Manual

Page 67

Advertising
background image

BM series

65

Entering the density of a liquid

Two ways to set the density of a liquid are available in the function table, "Liquid density input (

ldin

): by

entering the water temperature or by entering the density directly.


1 Press

the

MODE

key as necessary to select the density mode.

Press and hold the

MODE

key to select to "water temperature" or "liquid density".

In the density mode : The unit is "g (gram)". The processing indicator ( ) blinks.
Response adjustment is off.
If you cancel the density mode, press the

MODE

key.

Entering the water temperature

( ldin 0 )

2 The water temperature currently set (unit:°C, factory setting : 25°C) is displayed.

Use the following keys to change the value.

RE-ZERO

(+)key ....The key to increase the temperature by one degree.

(0°C is displayed after 99°C)

MODE

(-)key..........The key to decrease the temperature by one degree.

(99°C is displayed after 0°C)

PRINT

key............The key to store new water temperature, display

end

and return to the

density mode. Proceed to Step 1.

CAL

key...............The key to cancel the change and return to the density mode.

Proceed to Step 1.

The relation between the water temperature and density

°C

+0 +1 +2 +3 +4 +5 +6 +7 +8 +9

0

0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978

10

0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841

20

0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594

30

0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259

40

0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849

50

0.98804 0.98758 0.98712 0.98665 0.98618 0.98570 0.98521 0.98471 0.98422 0.98371

60

0.98320 0.98268 0.98216 0.98163 0.98110 0.98055 0.98001 0.97946 0.97890 0.97834

70

0.97777 0.97720 0.97662 0.97603 0.97544 0.97485 0.97425 0.97364 0.97303 0.97242

80

0.97180 0.97117 0.97054 0.96991 0.96927 0.96862 0.96797 0.96731 0.96665 0.96600

90

0.96532 0.96465 0.96397 0.96328 0.96259 0.96190 0.96120 0.96050 0.95979 0.95906

g/cm

3

Entering the density directly

( ldin 1 )

The

density

currently set (unit : g / cm

3

, factory setting : 1.0000g / cm

3

) is displayed.

Use the following keys to change the value.
The range to set the density is 0.0000g / cm

3

to 1.9999g / cm

3

.

RE-ZERO

1

(+)key....The key to set the value of the digit selected.

RANGE

1

(-)key........The key to select the digit to change the value.

PRINT

key............The key to store the change, display

end

and return to the density mode.

Proceed to Step 1.

CAL

key...............The key to cancel the change and return to the density mode. Proceed to Step 1.