12. Data Memory

Data memory is a function to store weighing data and calibration data in memory. The data stored in

memory are available for outputting at one time to a printer or personal computer.

Excluding date and time

Up to 200 sets

Weighing data

Including date and time

Up to 100 sets

Calibration report
Internal

calibration

External calibration

Calibration test report

Internal test calibration

External test calibration

Last 50 sets



12.1.

Notes On Using Data Memory

To use the memory function, set the "Data memory (

data

)" parameter

and "Time/Date output (

5-td

)" parameter of the function table. Refer to

"

10. The Function Table" for details on setting the data memory.

For weighing data, the data contents to be stored and the storage

capacity depend on the "Time/Date output (

5-td

)" parameter setting.

If a different type of data exists in memory,

"Clr"

blinks in the upper left

of the display.
For example: You want to store weighing data but calibration data or
unit mass data remains in memory.

"Clr"

blinks.

If the stored data is incomplete or deleted,

"err"

blinks in the upper left

of the display.
Under such a condition, before storing data, delete the data in memory
as follows:


Releasing "Clr" or "err"

1 Press and hold the

PRINT

key until

SClr "n"o

with

"no"

blinking is displayed, then release the key.


2 Press

the

RE-ZERO

key to display

Clr "g"o

with

"go"

blinking.

The type of data stored in memory appears in the upper left of the display as shown below:

Weighing data without time and date

-d-

Weighing data with time and date

d-t

Calibration report

Hi5


3 Press

the

PRINT

key to delete all the data in memory.


4 The

balance

displays

end

and returns to the weighing mode.

Upper left of

the display