Automatic self calibration, One-touch calibration – A&D Weighing BM-20 User Manual

Page 34

Advertising
background image

32

BM

series

8.2.

Automatic Self Calibration

(For variation of ambient temperature)

Automatic self calibration due to changes in temperature

This

function

automatically calibrates the balance when the balance detects an ambient

temperature change. If GLP output is selected in the function table, the balance outputs the
calibration report or stores the data in memory. Automatic self calibration functions even if the
display is turned off (standby state). Refer to "

9.1. Permit Or Inhibit" for the operation.

Caution

Place nothing on the weighing pan during automatic self calibration.
If something is on the weighing pan, the balance decides that it is in use and does not

perform automatic self calibration.

When weighing a light sample, using a long term weighing or installing the balance in a

system, turn off automatic self calibration.

Note

When turning on the balance with nothing on the pan, if a sample heavier than 0.5 g is
placed on the pan, the balance detects the state that a sample is placed on the pan and
does not perform the automatic self calibration.

The

mark

is " prior notice indicator of automatic self calibration".

When the balance detects a change in ambient temperature, this indicator blinks
and automatic self calibration is required. If the balance is not used for several
minutes with this indicator blinking, the balance performs automatic self calibration.
The environment will affect the time that the indicator blinks.

The balance is measuring calibration data. Do not allow vibration or drafts to affect
the balance while this indicator is displayed. After calibration, the balance returns
to indicate the previous display.

Advise
The balance can be used while the indicator blinks. But, it is recommended that to maintain the

accuracy, stop using the balance and confirm that there is nothing on the pan and allow the
balance to perform self calibration.


8.3.

One-Touch Calibration

(For Common Use)

This function calibrates the balance using the internal mass.
The only operation required is to press the

CAL

key.

1 Connect the AC adapter and warm up the balance for at

least one hour with nothing on the weighing pan.

2 Press

the

CAL

key to display

Cal in

.

3 The balance performs calibration using the internal mass.

Do not allow vibration or drafts to affect the balance.

4

If

GLP output is set,

glp

is displayed, the calibration test report

is output to the RS-232C interface and is stored in memory. Refer
to "GLP output (

info

)" and "Data memory (

data

)" of the function

table, "

11.2. GLP Report" and "12. Data Memory".

end

is displayed after the calibration.

5 The balance will automatically return to the weighing mode after calibration.