Calibration test using the internal mass – A&D Weighing BM-20 User Manual

Page 35

Advertising
background image

BM series

33

8.4.

Calibration Test Using the Internal Mass

This function tests the balance accuracy using the internal mass.

Calibration test does not perform calibration.
When GLP output is set, the calibration test report is output or stored.


1 Connect the AC adapter and warm up the balance at least

one hour.


2 Press and hold the

CAL

key until

CC in

is

displayed,

then release the key.



3 The balance measures the zero point.
Prevent vibration and drafts to affect the balance.

4 The

measured

zero point data is displayed.



5 The balance measures the internal mass.
Prevent vibration and drafts to affect the balance.

6 The value of the internal mass is displayed.

The normal range of the value is as follows:

Model

The internal mass The normal range

BM-20, BM-22

20.00000 g

±0.02 mg

BM-252, BM-200

BM-300, BM-500

200.0000 g

±0.2 mg


7

If

GLP output is set,

glp

is displayed, the calibration test

report is output to the RS-232C interface and is stored in
memory. Refer to "GLP output (

info

)" and "Data memory

(

data

)" of the function table, "

11.2. GLP Report" and "12.

Data Memory".

end

is displayed after the calibration.


9 The balance automatically returns to the weighing mode.