Error messages – HEIDENHAIN ND 920 User Manual

Page 38

Advertising
background image

38

Message

Problem

OFFSET DELETED

Offset compensation values for
encoder signals erased.

PARAM. ERASED

Check the operating parameters.
If this error recurs, contact your
service agency.

PGM ERASED

The program has been deleted.
If this error recurs, contact your
service agency.

PGM TOO LARGE

The maximum program length is
99 blocks.

PRESET ERASED

The datum points have been
erased. If this error recurs, contact
your service agency.

KEY W/O FUNCTION

This key currently has no
function.

TEMP. EXCEEDED

The temperature of the ND is too
high.

To clear error messages
When you have removed the cause of the error,
➤ press the CL key.

Error Messages

Message

Problem

CALL IS TOO FAST

Two commands for measured
value output occurred too close
together.

AMPL. X TOO LOW

The encoder signal is too weak.

The scale may be contaminated.

PROBING ERROR

The axis must move at least 6 µm
before probing.

DSR SIG. MISSING

The attached device is not
sending a DSR signal.....

INPUT ERROR

The entered value is not within
the permissible input range.

ERROR: REF. X

The spacing of the reference
marks as defined in P43 is not
the same as the actual spacing.

FORMAT ERROR

Data format, baud rate, etc., do
not agree.

FRQ. EXCEEDED X

The input frequency for this encoder

input is too high. This can occur
when the scale is moved too fast.

COMP. DELETED

Compensation values for non-
linear axis error compensation
erased.

Error Messages