Operating Parameters

P40.1 to P40.3

Select type of axis error compensation

No axis error compensation

AXIS COMP X OFF

Linear error compensation active,
multipoint error comp. not active

AXIS COMP X LIN

Multipoint error compensation active,
linear error compensation not active

AXIS COMP X F(a)

P41.1 to P41.3

Linear axis error compensation

Input range (µm):

99999 to +99999

Factory setting:

0

Example

Displayed length L

d

= 620.000 mm

Actual length (as determined for example with
the VM 101 from HEIDENHAIN)
L

a

= 619.876 mm

Difference DL = L

a

– L

d

= –124 µm

Compensation factor k:
k = DL/L

d

= –124 µm/0.62 m = –200 [µm/m]

P43.1 to P43.3

Reference marks

One reference mark

0

Distance-coded with 500 x SP

500

Distance-coded with 1000 x SP

1000

Distance-coded with 2000 x SP

2000

Distance-coded with 5000 x SP

5000

(SP = signal period)

P44.1 to P44.3

Reference mark evaluation

Reference mark evaluation active

REF. MODE X ON

Not active

REF. MODE X OFF

P45.1 to P45.3

Encoder monitoring

Amplitude and frequency
monitoring active

ALARM X ON

Not active

ALARM X OFF

P48.1 to P48.3 Activate axis display

Axis display active

AXIS DISPL.X ON

Not active

AXIS DISPL.X OFF

P49.1 to P49.3 Axis designation, measured value output

2)

For output of the measured value, an axis designation can be
defined with the number of the ASCII character. The axis
designation is output together with the measured value.

Input range:

0 to 127

Measured value output disabled

0

ASCII character from ASCII table

1 to 127

Factory setting:

P49.1

88

P49.2

89

P49.3

90

P50 Baud rate

1) 2)

110 / 150 / 300 / 600 / 1200 / 2400 / 4800 / 9600 /
19 200 / 38 400

P51 Blank lines for measured value output

1) 2)

Input range:

0 to 99

Factory setting

1

1)

User parameter

2)

Only on units with RS-232-C/V.24 and EXT connection