Measured value output – HEIDENHAIN ND 920 User Manual

Page 58

Advertising
background image

58

Pin 1(0V)

Pin 9

Pin 8

Pin 1(0V)

EXT(X41)

EXT(X41)

Measured value output over the Contact and Pulse inputs

Measured value output over the Contact input (pin 9 on X41) and Pulse
input (pin 8 on X41) can be triggered when these inputs are closed
against 0 V.

The measured values are output over the TXD line of the RS-232-C
interface.

A commercially available switch can be attached to the Contact input.
This switch generates a signal for data output when it makes contact
against 0 V.

The Pulse input can be triggered with TTL logic devices (for example,
SN74LSXX).

Delay times for data output

Latch signal duration: Pulse

t

e

1.2 µs

Latch signal duration: Contact

t

e

7 ms

Storage delay: Pulse

t

1

0.8 µs

Storage delay: Contact

t

1

4.5 ms

Data output after

t

2

30 ms

Regeneration time

t

3

0

Duration of data output in seconds:

t

D

=

t2

t1

t3

tD

te

te

Next possible signal for measured value output:
t

E

= t

1

+ t

2

+ t

D

+ t

3

[s]

176 x number of axes + 11 x number of blank lines

Baud rate

Measured Value Output

Advertising