Measured value output – HEIDENHAIN ND 920 User Manual

Page 55

Advertising
background image

55

Example of measured value output:

Parameter settings:

P49.1 =

88

(“X”)

P49.2 =

89

(“Y”)

P49.3 =

90

(“Z”)

P51

=

0

(no blank lines)

P97

=

69

(“E”)

Output:

E (CR)(LF)
X=...(CR)(LF)
Y=...(CR)(LF)
Z=...(CR)(LF)

Measured Value Output

Measured values can be output over the RS-232-C/V.24 interface. This
can be done with the following functions:

Probing with the KT Edge Finder
“Contact” input on X41
“Pulse” input on X41
CTRL B over the RS-232-C interface
Measured value output with the HOLD POS key.

Parameter P23 influences the display mode for measured value output.
It is not effective, however, for output from probing.

Code letters with the measured value

Parameter P97 allows you to select a code letter to be output together
with the measured value when using Probe, Contact or Pulse. The
decimal number you enter in the parameter is the ASCII character
number in the ASCII table. If you enter 0, no code letter will be output.
The code letter enables you to recognize whether the measured value
was generated with CTRL B or with an external signal.

Axis designation for measured value output

Parameter P49 allows you to enter an axis designation for each
measured value that is output. The decimal number you enter in the
parameter is the ASCII character number in the ASCII table. If you enter
0, no axis designation will be output.