Linear encoders, Setting the display step, Display step, signal period and subdivision – HEIDENHAIN ND 920 User Manual

Page 48

Advertising
background image

48

Linear Encoders

Linear Encoders

Setting the display step with linear encoders

The display step depends on the
•

signal period of the encoder (P31) and the

•

subdivision (P32).

Both parameters are entered separately for each
axis.

For linear measurement using nut/ballscrew
arrangements and rotary encoders, calculate the
signal period as follows:

Signal period [µm] =

Drivescrew pitch [mm] x 1000

Line count

Display step, signal period and subdivision for linear encoders

Display step

P31: Signal period [µm]

2 4 10 20 40 100

2 4 10 20 40 100

2 4 10 20 40 100

2 4 10 20 40 100

2 4 10 20 40 100 200

200

200

200

200

12

12

12

12

12 800

800

800

800

800

[mm]

[inches]

P32: Subdivision

0.000 02 0.000 001

100

–

–

–

–

–

–

–

0.000 05 0.000 002

40

80

–

–

–

–

–

–

0.000 1

0.000 005

20

40

100

–

–

–

–

–

0.000 2

0.000 01

10

20

50 100

–

–

–

–

0.000 5

0.000 02

4

8

20

40

80

–

–

–

0.001

0.000 05

2

4

10

20

40

100

–

–

0.002

0.000 1

1

2

5

10

20

50

100

–

0.005

0.000 2

0.4

0.8

2

4

8

20

40

–

0.01

0.000 5

0.2

0.4

1

2

4

10

20

–

0.02

0.001

–

–

0.5

1

2

5

10

–

0.05

0.002

–

–

0.2

0.4

0.8

2

4

–

0.1

0.005

–

–

0.1

0.2

0.4

1

2

128

0.2

0.01

–

–

–

–

–

–

–

64

Advertising