8 counter function – HEIDENHAIN PWM 20 User Manual
Page 204

November 2014
Checking incremental encoders
207
6.3.8 Counter function
The "Counter" function serves to check the counting of encoders.
Signal periods are counted and displayed. The function starts immediately after you select
"Counter".
The Counter characteristics field contains:
Position
=
Display value in signal periods
Reference =Determined distance between two reference marks
Direction =
Display of traverse direction or direction of rotation (in tabular view:
positive
and negative
, the current position is marked by an asterisk
*.)
Trigger
=When the reference mark is traversed, the LED color changes to "Green“ (pulse).
Note
The "Position" column contains the signal periods.
Freeze the counter screen
Reset Counter and Position to zero
Every reference mark clears Counter and Position
Clear Counter and Position and restart with the next reference mark
Change of counting directions for Counter and Position
Clear the table and the graphics
Return to basic menu (Connect/disconnect encoder)