Video Test Generator User Guide

449

NCPP

Number Clocks Per Pixel

Class

Format parameter setting - Digital video signal

Description

Specifies the number of clocks per pixel (double-clocking factor for whole line). Some

industry standard formats (e.g. 480i) have pixel rates that are below the minimum pixel

rate restriction (of 25 MHz) required by the TMDS interface. In these cases, a feature

known as “double-clocking” is used to raise the TMDS clock rate to an acceptable

frequency.

Double-clocking is controlled by the NCPP format parameter. When NCPP is set to 1

(normal single-clocked), each pixel, output by the generator, is paired with one clock

pulse. In this case, the pixel rate is equal to the TMDS clock rate. By setting NCPP to 2

(double-clocked), two clocks are send during each pixel period and the TMDS clock rate is

raised to 2 times the pixel rate.

Note that double clocking with NCPP and pixel repetition (NPPP) cannot be applied at the

same time.

Command syntax

NCPP factor

factor

1 or 2

Related commands

NBPA

,

NPPP

Example

NCPP 2
FMTU